o SERIAL MODBUS configure serial output,
o PASSWD Enter 4 character password to protect access to MENU Level, CAL Menu, CONFIG
Menu and SIM Menu (simulate). Each level can be turned ON or OFF and can have a unique
password.
 SENSOR enters the sensor configuration menu.
o Choose SENSOR 1 or 2
o COMP Dual Channel Only, Sets compensation type: Dissociation (pKa),
+
NH
, Free Chlorine, HF, S
4
Sensor 1, Percentage % change per pH, Total Chlorine -5%.
o Qty of SENSORS, Choose 1 or 2
 Load Default resets all Menus to factory default configuration.
 Dampen, sets the number of measurements averaged for the displayed PV
3.2.4 INFO (I
NFORMATION
The Information Menu provides two choices,
 Transmitter Screen, details the Name, Power type, Serial #, Firmware version
and the output configuration(s).
 Sensor Screen, details the Name, Part #, Serial # and three sets of Calibration data.
 COMP, displays the pKa, the sensor affected and the dissociation Factor,
3.2.5 SIM (S
IMULATION
The Simulation menu allows the Input or Output signals to be simulated.
Page 25
o Disable, Inactivates relay and removes the
relay button from the HOME Screen display.
HOLD, Freezes outputs at current value and locks relays in
their current state.
Hold Timeout, Removes HOLD after a certain period
of time, default setting: No Timeout, selections include 15 minutes, ½ hour, 1
hour
ADDRESS, enter address: 001 to 247
BAUD, Choose baud rate, default 9600
FORMAT, set serial data format, default value: 8N1, 8 bit, no
parity bit, 1 stop bit
MENU
ON/OFF _ _ _ _ Locks Main Menu
CAL
ON/OFF _ _ _ _ Locks CAL and CONFIG
CONFIG
ON/OFF _ _ _ _ Locks CONFIG
SIM
ON/OFF _ _ _ _ Locks SIM and CONFIG
TYPE, Allows T80 transmitter to configure the S80 sensor. For
use only when switching the measurement electrode type in
an S80 sensor, i.e. for a pH electrode to a pION electrode.
Select Sensor Type: pH, ORP, DO
Conductivity, Resistivity, Cu
T COMP, Enter % temperature compensation per degree: pH,
0.33%, Total Chlorine 0%,
-2
, Interference, X ppm Sensor 2 = 1 ppm
M
)
ENU
M
)
ENU
+
-
, NH
, NH
, Br
, Ca
2
3
4
++
-
-
-
+
+
, CN
, F
, NO
, K
, Ag
, Na
3
++
-
, Cl
,
+
--
, S
Model TC80