Ph Calibration Procedures; Auto Cal Using Ph 4.01, 7.00, 10.00 Buffers; Auto Cal Using Other Ph Buffers; Standardize - ECD TC80 Instruction Manual

Total chlorine analyzer; dechlorination analyzer
Table of Contents

Advertisement

4.1
H C
P
P
ALIBRATION
AUTO Calibration recognizes pH 4.01, pH 7.00 and pH 10.00 buffer solutions for automatic, temperature
compensated calibrations. Any calibration solutions can be used but the pH value will have to be entered
manually. Follow the steps below to accomplish a pH calibration. Example uses pH 7.00 and pH 4.01 buffers.
4.1.1 AUTO C
AL USING P
Action
Double Press any Button
Press HOLD
Press CAL
Press Yes/NO
Press AUTO then CAL 1
Press YES
Press YES
Press YES
Press OK
Press HOME
Press HOLD
Press EXIT
4.1.2 AUTO C
AL USING OTHER P
Action
Double Press any Button
Press HOLD
Press CAL
Press YES/NO
Press AUTO then CAL 1
Press NO
Press ▲ ▼ NEXT
Press OK
Press YES
Press YES
Press NO
Press ▲ ▼ NEXT
Press OK
Press YES
Press OK
Press HOME
Press HOLD
Press EXIT
4.1.3 S
TANDARDIZE
Leave the sensor in the process solution, take a grab sample from the process and determine the pH or place
sensor in a calibration standard solution.
Action
Double Press any Button
Press HOLD
Page 33
ROCEDURES
H 4.01, 7.00, 10.00
Prompt
MENU HOME, Hold is OFF
Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration
Is this a new Sensor? (Yes erases CAL Log in INFO, NO adds CAL to existing Log)
Place Sensor in CAL Solution ( use pH 7.00 buffer)
STABILIZING, 7.00 pH xx.x mV, 7.00 pH corrected Accept Cal 1?
CAL1 Value 7.00 pH, Continue to CAL2? Move sensor to 4.01 pH buffer solution
STABILIZING, 4.00 pH xxx.x mV, 4.00 pH corrected Accept Cal?
OFFSET: 7.00 pH xx.x mV, SLOPE: 59.16 mV/pH (data written to Log)
Calibration complete
Hold is ON
Turn off Hold
Main Display
H
BUFFERS
Prompt
MENU HOME, Hold is OFF
Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration
Is this a new Sensor? (Yes erases CAL Log in INFO, NO adds CAL to existing Log)
Place Sensor in CAL Solution
STABILIZING, 6.86 pH 8.2 mV, 7.00 pH corrected Accept Cal?
Enter CAL 1 Value
6.86 pH (use arrows and NEXT to enter pH Buffer value)
6.86 pH, 8.2 mV, Accept this Value
CAL 1 Value 6.86 pH, Continue to CAL 2? (Place Sensor in 2
STABILIZING, 9.18pH 135.6 mV, 10.00 pH corrected Accept Cal?
Enter CAL 2 Value
9.18 pH (use arrows and NEXT to enter pH Buffer value)
9.18 pH, 135.6 mV, Accept this Value
OFFSET: 6.86 pH 8.2 mV, SLOPE: 59.16 mV/pH (data written to Log)
Calibration complete
Hold is ON
Turn off Hold
Main Display
Prompt
MENU HOME, Hold is OFF
Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration
BUFFERS
nd
calibration buffer)
Model TC80