1.1 F
EATURES
Separate, economical, easily replaceable Nitrate, Chloride, and optional pH electrodes
Fast and Accurate Nitrate Measurement NO
Automatic compensation for Cl interference
Temperature compensated NO
Rugged PVC design with removable electrode guard for easy maintenance
Integral Spray Head Cleaner
o Cleans sensor in situ with turbulence caused by pressurized air
Internal Signal Conditioning allows up to 200 meters between Sensor and Analyzer
1.2 HYDRA-DS S
ENSOR
Three Electrode System with spray cleaner
Nitrate ISE (NO
-N) is the primary measurement
3
Chloride ISE is used for compensation of the NO
Optional pH or NH
+
4
The Sensor is waterproof with an ingress rating of IP68
1.2.1 M
R
EASUREMENT
ANGE
NO
0.1 to 1000 ppm
-N
3
Cl
0.1 to 10,000 ppm
pH
2-12 ph
NH
+
0.1 to 1000 ppm
4
1.2.2 O
T
PERATING
EMPERATURE
0 °C to 40 °C (32 °F to 104 °F)
1.2.3 M
/M
F
R
IN
AX
LOW
ATE BY THE SENSOR
Minimum 0.1 m/s
Maximum 3.0 m/s
1.2.4 W
M
ETTED
ATERIALS
PVC, PEC, PVDF, PTFE, Viton, Glass, 316 SS
1.2.5 A
CCURACY
± 3% of reading, dependent on Calibration
1.2.6 R
T
ESPONSE
IME
T90 approximately 1 minute
1.2.7 E
L
LECTRODE
IFE
ISEs: 4-6 months, typical
pH electrode: 6-12 months, typical
HYDRA-DS NO
-N
3
3
-N measurement
3
S
PECIFICATIONS
electrode
or NO
-N
3
signal
3
Page 11