General errors

*
Cause
– Printer switched off
– No mains power supply
connected
– Interface cable connected
– No paper available
– AutoStore function is switched
on with a longer interval duration
(Int 1)
Cause
– AutoStore function
(Int 1) or Data transmission (Int
2) is switched on
f
Cause
– Printer is switched on
– Interface cable connected
Cause
– Paper inserted with wrong side
upwards
*
only on the inoLab pH/ION 735 with integrated printer
inoLab pH/ION 735
Remedy
– Switch on printer
– Connect mains power
supply
– Unplug cable
– Replace roll of paper
– Switch off AutoStore
function
Remedy
– Switch off functions
Remedy
– Switch off printer
– Unplug cable
Remedy
– Turn the roll of paper around
and insert it with the other
side upwards