To make automatic measurements

You can use automatic measurements on any channel source or any running
math function. Cursors are turned on to focus on the most recently selected
measurement (right-most on the measurement line above the softkeys on the
display).
1 Press the
Quick Meas
2 Press the
Source
softkey to select the channel or running math function
on which the quick measurements will be made.
Only channels or math functions that are displayed are available for
measurements. If a portion of the waveform required for a measurement is not
displayed or does not display enough resolution to make the measurement, the
result will be displayed with a message such as greater than a value, less than
a value, not enough edges, not enough amplitude, incomplete, or waveform is
clipped to indicate that the measurement may not be reliable.
3 Press the
Clear Meas
the measurement results from the measurement line above the
softkeys.
When Quick Meas is pressed again, the default measurements on an analog
channel will be will be Frequency and Peak-Peak.
4 Choose what measurements you want on that source by pressing the
softkey.
5 To turn off
Quick Meas
illuminated.
For detailed information about making automatic measurements, refer to the
"Making Measurements" chapter.
key to display the automatic measurement menu.
softkey to stop making measurements and to erase
, press the
Quick Meas
Front-Panel Overview
To make automatic measurements
key again until it is not
2- 23