Making Measurements

To use the Roll horizontal mode

4 Characterize the glitch with delayed sweep.
To characterize the glitch with delayed sweep, follow these steps:
• Press the Main/Delayed key, then press the Delayed softkey.
• To obtain a better resolution of the glitch, expand the time base.
Use the horizontal delay time knob (
)to pan through the waveform to set the
expanded portion of the main sweep around the glitch.
To characterize the glitch, use the cursors or the automatic measurement
capabilities of the oscilloscope.
To use the Roll horizontal mode
• Press the
key, then press the
softkey.
Main/Delayed
Roll
Roll mode causes the waveform to move slowly across the screen from right to
left. It only operates on time base settings of 500 ms/div and slower. If the
current time base setting is faster than the 500 ms/div limit, it will be set to
500 ms/div when Roll mode is entered.
In Normal horizontal mode, signal events occurring before the trigger are
plotted to the left of the trigger point ( ) and signal events after the trigger
plotted to the right of the trigger point.
In Roll mode there is no trigger. The fixed reference point on the screen is the
right edge of the screen and refers to the current moment in time. Events that
have occurred are scrolled to the left of the reference point. Since there is no
trigger, no pretrigger information is available.
If you would like to clear the display and restart an acquisition in Roll mode,
press the Single key.
Use Roll mode on low-frequency waveforms to give a display much like a strip
chart recorder. It allows the waveform to roll across the display.