U3045AM04
Electrostatic Discharge Protection
Electrostatic Discharge Protection
Protection against electrostatic discharge (ESD) is essential while removing
assemblies from or connecting cables to the instrument. Static electricity can
build up on your body and can easily damage sensitive internal circuit
elements when discharged. Static discharges too small to be felt can cause
permanent damage. To prevent damage to the instrument:
— al ways have a grounded, conductive table mat in front of your test
— al ways wear a grounded wrist strap with grounding cord, connected to a
— al ways wear a heel strap (9300-1126) when working in an area with a
— al ways ground yourself before you clean, inspect, or make a connection to
— al ways ground the center conductor of a test cable before making a
Keysight U3045AM04 User's and Service Guide
equipment.
grounded conductive table mat, having a 1 MΩ resistor in series with it,
when handling components and assemblies or when making connections.
conductive floor. If you are uncertain about the conductivity of your floor,
wear a heel strap.
a static-sensitive device or test port. You can, for example, grasp the
grounded outer shell of the test port or cable connector briefly.
connection to the analyzer test port or other static-sensitive device. This
can be done as follows:
1. Connect a short to one end of the cable to short the center conductor
to the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the
cable connector.
3. Connect the other end of the cable to the test port and remove the
short from the cable.