Electromagnetic Compatibility - Advan AMM215WTTP User Manual

Lcd color display
Table of Contents

Advertisement

Electromagnetic Compatibility

Like other electrical medical equipment, the AMM215WTTP requires special precautions to
ensure electromagnetic compatibility with other electrical medical devices. To ensure
electromagnetic compatibility (EMC), the AMM215WTTP must be installed and operated
according to the EMC information provided in this manual.
Note
The AMM215WTTP has been designed and tested to comply with
EN 60601-1-2 requirements for EMC with other devices.
Caution
Portable and mobile RF communications equipment may affect the
normal function of the AMM215WTTP.
Warning
Do not use cables or accessories other than those provided with
the AMM215WTTP, as this may result in increased electromagnetic emissions
or decreased immunity to such emissions.
Warning
If the AMM215WTTP is used adjacent to or stacked with other
equipment, observe and verify normal operation of the AMM215WTTP in the
configuration in which it will be used prior to using it in a surgical procedure.
Consult the tables below for guidance in placing the AMM215WTTP.
Manufacturer's Declaration - Electromagnetic Emission
The Model AMM215WTTP is intended for use in the electromagnetic environment specified below.
The user of AMM215WTTP should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage Fluctuations
IEC61000-3-3
Manufacturer's Declaration - Electromagnetic Immunity
The Model AMM215WTTP is intended for use in the electromagnetic environment specified below.
The user of AMM215WTTP should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC61000-4-2
Electrical fast transient
/burst IEC61000-4-4
Surge
IEC61000-4-5
Compliance
Electromagnetic Environment - Guidance
The Model AMM215WTTP uses RF energy only for its internal
Group 1
function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment
Class B
AMM215WTTP is suitable for use in all establishments, including
Class D
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
Complies
domestic purposes
IEC 60601 Test Level
6 kV contact
8 kV air
2 kV for power supply lines
1 kV for input / output lines
1 kV differential mode
2 kV common mode
Compliance Level
6 kV contact
8 kV air
2 kV for power supply lines
1 kV for input / output lines
1 kV differential mode
2 kV common mode
7
Electromagnetic
Environment - Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relativehumidity should be
at least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment
Mains power quality
should be that of a typical
commercial or
hospital environment.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents