Electromagnetic Compatibility

Like other electrical medical equipment, the AMM190WTD2 requires special precautions
to ensure electromagnetic compatibility with other electrical medical devices. To ensure
electromagnetic compatibility (EMC), the AMM190WTD2 must be installed and operated
according to the EMC information provided in this manual.
Note
The AMM190WTD2 has been designed and tested to comply with IEC
60601-1-2:2001 requirements for EMC with other devices.
Caution Portable and mobile RF communications equipment may affect
the normal function of the AMM190WTD2.
Warning Do not use cables or accessories other than those provided
with the AMM190WTD2, as this may result in increased electromagnetic emis-
sions or decreased immunity to such emissions.
Warning
If the AMM190WTD2 is used adjacent to or stacked with other
equipment, observe and verify normal operation of the AMM190WTD2 in the
configuration in which it will be used prior to using it in a surgical procedure.
Consult the tables below for guidance in placing the AMM190WTD2.
Manufacturer's declaration - electromagnetic emission
The Model AMM190WTD2 is intended for use in the electromagnetic environment specified below.
The customer or the user of AMM190WTD2 should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage Fluctuations
IEC61000-3-3
Manufacturer's declaration - electromagnetic immunity
The Model AMM190WTD2 is intended for use in the electromagnetic environment specified below.
The customer or the user of AMM190WTD2 should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Electrostatic discharge
6 kV contact
(ESD)
8 kV air
IEC61000-4-2
Electrical fast transient
2 kV for power supply lines
/burst IEC61000-4-4
1 kV for input / output lines
Surge
1 kV differential mode
IEC61000-4-5
2 kV common mode
Compliance
Electromagnetic environment - guidance
The Model AMM190WTD2 uses RF energy only for its internal
Group 1
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment
Class B
AMM190WTD2 is suitable for use in all establishments, including
Class D
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes
Complies
Compliance Level
6 kV contact
8 kV air
2 kV for power supply lines
1 kV for input / output lines
1 kV differential mode
2 kV common mode
AMM190WTD2 29
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.