Contents Warnings and Cautions ............ 5 Symbol Definitions .............. 8 Product Description ............9 Intended Use ................9 Indications/Contraindications ..........10 Package Contents ..............11 Device Features ..............12 Setup and Interconnection .......... 14 Connection Ports ..............14 Connecting AC Power ............15 Basic Video Setup ..............16 Operating the Display .............
Warnings and Cautions Please read this manual and follow its instructions carefully. The words warning, caution, and note carry special meanings and should be carefully reviewed: Warning Indicates risks to the safety of the patient or user. Failure to follow warnings may result in injury to the patient or user. Caution Indicates risks to the equipment. Failure to follow cautions may result in product damage. Note Provides special information to clarify instructions or present additional useful information.
Page 6
Use only the proprietary Vision Elect WHDTV power supply (P/N 240- 030-90, Manufacturer: JEC Korea, Model No: JMW110KA2400F07) for the Vision Elect WHDTV display (model 240-030-970). Completely Connect equipment to a receptacle labeled “Hospital Only” or “Hospital Grade” to achieve grounding reliability.
Page 7
(model 240-030-970) display must be installed and operated according to the EMC information provided in this manual. The VISION ELECT WHDTV (model 240-030-970) display has been tested under the UL 60601-1 standard and is UL listed for medical application. The warranty is void if any of these warnings or cautions are disregarded.
Symbol Definitions Dangerous High Voltage Consult accompanying documents Direct Current Protective earth ground For indoor use only DC power control switch Serial Number Top - Bottom Fragile Do not get wet Maximum Stacking Manufacturer European Authorized Representative Indicates proof of conformity to applicable European Economic Community journal of the European Communities Medical Equipment is in accordance with UL 60601-1 and CAN/CSA C22.2 Tested to comply with FCC Class B standards...
Product Description The Vision Elect WHDTV (VE WHDTV) is a high-definition, widescreen LCD surgical display with a maximum resolution of WUXGA (1920 × 1200 at 60 Hz). The VE WHDTV supports various video inputs, including digital RGB, analog RGB, serial digital interface (SDI), component video (YPbPr/RGB), S-video, and C-video.
Indications/Contraindications The VE WHDTV is indicated for use as an accessory to an endoscopic surgical camera during general surgery, general laparoscopy, nasopharyngoscopy, ear endoscopy, sinusoscopy, and plastic surgery wherever a laparoscope/endoscope/ arthroscope is indicated for use. Some of the more common endoscopic surgeries where the VE WHDTV is indicated for use include: cholecystectomy;...
Package Contents a. 1 DVI cable b. 1 VGA HDDB1 cable c. 1 AC adapter (Stryker P/N 240-030-90, Manufacturer: JEC, Model No: JMW110KA2400F07) d. 1 Hospital-grade AC power cord e. 4 M4 × 10mm VESA screws 1 BNC cable g. 1 S-Video cable h.
Device Features Front panel Operate the display using the rotary control located on the front panel. A list of the display controls and their functions is provided below. 1. Power LED Shines green if the display is powered on; blinks red if the display is in standby mode. 2. Power switch (soft) Powers the display on and off 3. Rotary control...
Page 13
Rear panel 1. Power connector Connects to the 24V DC power converter 2. Power switch (hard) Powers on and off the input DC power 3. VESA mounting Provide access points for mounting the monitor holes (100mm) 4. Connector tags Indicate the types of video connectors. 5. Cable-management Organize cables clamps...
Setup and Interconnection Stryker Endoscopy considers instructional training, or inservice, an integral part of the transmitter. Your local Stryker Endoscopy sales representative will perform at least one inservice at your convenience to help set up your equipment and instruct you and your staff on its operation and maintenance. To schedule an inservice, contact your local Stryker Endoscopy representative after your equipment has arrived.
Connecting AC Power Connect the AC power, using the supplied hospital-grade power cord. Connect the power cord to the power supply. (240-030-90, manufacturer: JEC Korea, model JMW110KA2400F07) Connect the power supply to the 24V input on the display.
Basic Video Setup VE WHDTV SDC or SIDNE® VCR or Printer Camera Route the video input 1 from the camera to the SDC or SIDNE. Route the video input 1 to the DVI input on the display. Connect the C-video input on the display to the C-video output on a VCR or printer.
Operating the Display Accessing the On-Screen Display Operate the display using the rotary control located on the front panel. A list of the display controls and their functions is provided below. Power LED Rotary Control Power Switch (Soft) Power LED: Indicates menu current status. Displays green if display is powered on;...
Using the On-Screen Display The display provides an on-screen display to help navigate through the various display-adjustment menus. Press the Rotary Control to activate the on-screen display (OSD) menu. Rotate the Rotary Control to move up or down through the menu. The parameter will be highlighted when selected.
On-Screen Display Menus Item Function Description Range Specialty Color Temperature * Chooses between color temperatures for Standard, Arth, Lap, PACS, or Norm Red balance -128 to 128 Green Green balance -128 to 128 Blue Blue balance -128 to 128 Gamma Gamma value 0.1 to 2., S0, S1, S2...
Page 20
Item Function Description Range Auto Source Select Adjusts Auto Source Select between on and off Restore Factory Sets to factory default Settings Key lock Sets to Key lock mode Wireless Mac ID Unique Machine ID of WHDTV Transmitter Status Status Message Information User Name Entry Enters custom username display for boot-up display...
No specific liquid or chemical is necessary for cleaning the VISION ELECT WHDTV (model 240-030-970) LCD display. Use only non-abrasive cloths and cleaning solutions to clean similar equipment used in hospitals. Clean the plastic areas of the display with a dry soft cloth, or a soft cloth lightly moistened with mild detergent solution.
Troubleshooting Before returning your LCD display for service, consult the troubleshooting list below: Problem Current Status Remedy No picture LED on Using the OSD, adjust the brightness and contrast to maximum, or reset them to their default settings. LED off Check the power switch at the front and back of the display.
Electromagnetic Compatibility Like other electrical medical equipment, the VISION ELECT WHDTV (model 240-030-970) requires special precautions to ensure electromagnetic compatibility with other electrical medical devices. To ensure electromagnetic compatibility (EMC), the display must be installed and operated according to the EMC information provided in this manual.
Page 26
Guidance and Manufacturer’s Declaration: Electromagnetic Emissions The Model 240-030-970 is intended for use in the electromagnetic environment specified below. The customer or the user of Model 240-030-970 should ensure that it is used in such an environment.. Emissions test Compliance...
Page 27
Guidance and Manufacturer’s Declaration--Electromagnetic Immunity Model 240-030-970 is intended for use in the electromagnetic environment specified below. The customer or the user of Model 240-030-970 should ensure that it is used in such an environment. IEC 60601 Test Compliance Electromagnetic Environment--...
Page 28
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity Model 240-030-970 is intended for use in the electromagnetic environment specified below. The customer or the user of Model 240-030-970 should ensure that it is used in such an environment. IEC 60601 Test Compliance...
Page 29
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Model 240-030-970 System The Model 240-030-970 system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the Model 240-030-970 system can help...
Warranty Stryker Endoscopy warrants all products, subject to the exceptions provided herein, to be free from defects in design, materials and workmanship and to substantially conform to the product specifications contained in the documentation provided by Stryker Endoscopy with the products for a period of one year from the date of purchase (the “Warranty Period”).
Notwithstanding the above, the following products are warranted for a period of ninety (90) days from the date of purchase: Scopes, Fiber Optic Cables, VCRs, Displays, and Printers. Replacement light bulbs are warranted for a period of sixty (60) days from the date of purchase. TO THE FULLEST EXTENT PERMITTED BY LAW, THE EXPRESS WARRANTY SET FORTH HEREIN IS THE ONLY WARRANTY APPLICABLE TO THE PRODUCTS AND IS EXPRESSLY IN LIEU OF ANY OTHER...
Page 33
ADVAN Int’l Corp. 47817 Fremont Blvd. Fremont, CA 94538 USA 1-510-490-1005 2008/06 1000401089 A...