Electromagnetic Compatibility

Like other electrical medical equipment, the VISION ELECT WHDTV
(model 240-030-970) requires special precautions to ensure electromagnetic
compatibility with other electrical medical devices. To ensure electromagnetic
compatibility (EMC), the display must be installed and operated according to the
EMC information provided in this manual.
Note
Caution
Caution
Caution
Warning
25
32
The VISION ELECT WHDTV (models 240-030-970 and -971)
has been designed and tested to comply with IEC 60601-1-
2:2001 requirements for EMC with other devices.
Portable and mobile RF communications equipment may
affect the normal function of the display.
Do not use cables or accessories other than those
provided with the display, as this may result in increased
electromagnetic emissions or decreased immunity to such
emissions.
If the display is used adjacent to or stacked with other
equipment, observe and verify normal operation of the
display in the configuration in which it will be used prior to
using it in a surgical procedure. Consult the tables below for
guidance in placing the display.
When this device is connected with other electrical
equipment, leakage currents may be additive. To minimize
total leakage current per patient, ensure that all systems are
installed according to the requirements of IEC 60601-1-1.