Electromagnetic Compatibility
Like other electrical medical equipment, the
precautions to ensure electromagnetic compatibility with other electrical medical devices. To
ensure electromagnetic compatibility (EMC), the AMM213TD/AMM240WTD must be installed
and operated according to the EMC information provided in this manual.
Note
The AMM213TD/AMM240WTD has been designed and tested to comply with
IEC 60601-1-2:2001 requirements for EMC with other devices.
Caution Portable and mobile RF communications equipment may affect
the normal function of the AMM213TD/ AMM240WTD.
Warning Do not use cables or accessories other than those provided
with the AMM213TD/AMM240WTD, as this may result in increased electromag-
netic emissions or decreased immunity to such emissions.
Warning
If the AMM213TD/AMM240WTD is used adjacent to or stacked
with other equipment, observe and verify normal operation of the AMM213TD/
AMM240WTD in the configuration in which it will be used prior to using it in a
surgical procedure. Consult the tables below for guidance in placing the AM-
Manufacturer's declaration - electromagnetic emission
The Model AMM213TD/AMM240WTD is intended for use in the electromagnetic environment specified below.
The customer or the user of AMM213TD/AMM240WTD should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage Fluctuations
IEC61000-3-3
Manufacturer's declaration - electromagnetic immunity
The Model AMM213TD/AMM240WTD is intended for use in the electromagnetic environment specified below.
The customer or the user of AMM213TD/AMM240WTD should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC61000-4-2
Electrical fast transient
/burst IEC61000-4-4
Surge
IEC61000-4-5
Compliance
The Model AMM213TD/AMM240WTD uses RF energy only for
its internal function. Therefore, its RF emissions are very low
Group 1
and are not likely to cause any interference in nearby electronic
equipment
Class B
AMM213TD/AMM240WTD is suitable for use in all
Class D
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes
Complies
IEC 60601 Test Level
6 kV contact
8 kV air
2 kV for power supply lines
1 kV for input / output lines
1 kV differential mode
2 kV common mode
AMM213TD/AMM240WTD
Electromagnetic environment - guidance
Compliance Level
6 kV contact
8 kV air
2 kV for power supply lines
1 kV for input / output lines
1 kV differential mode
2 kV common mode
requires special
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.