Troubleshooting

Fault
❑ LED at the con-
denser is off, no fan
noise.
❑ LED is flashing
orange, no change
to continuous green
light.
❑ Continuous red
light at the con-
denser.
page 36 of 44
Possible cause
➨ Exhaust waste vapor
condenser Peltronic
not switched on or con-
troller switched off?
➨ Electrical power cord
not plugged in, electri-
cal supply failure?
➨ Exhaust waste vapor
condenser Peltronic
defective?
➨ Very high condensation
power?
➨ Ambient temperature
too high?
➨ Ambient temperature
much too high?
➨ Very high condensa-
tion power and ambient
temperature too high?
➨ Exhaust waste vapor
condenser Peltronic
defective?
Remedy
✔ Switch on condenser and/
or controller.
®
✔ Plug in power cord.
Check fuse.
✔ Contact local distributor.
®
✔ Check process parame-
ters (e.g., reduce pump-
ing speed of the vacuum
pump, increase cooling
at the vacuum side of the
system).
✔ Comply with maximum
permissible ambient tem-
perature.
✔ Comply with maximum
permissible ambient tem-
perature.
✔ Check process parame-
ters (e.g., reduce pump-
ing speed of the vacuum
pump, increase cooling
at the vacuum side of
the system). Comply with
maximum permissible
ambient temperature.
✔ Contact local distributor.
®