Make sure ventilation is adequate to maintain recommend-
NOTICE
ed operating temperature. Keep a minimum distance of
2 in (5 cm) between the cooling fan and surrounding items
(e.g., housing, walls, etc.), or else install an external auto-
matic ventilation system.
Check ventilation slots regularly for dust/dirt. Clean venti-
lation slots if necessary to avoid a reduction of ventilation.
Allow the equipment to equilibrate to ambient temperature
if you bring it from cold environment into a room prior to
operation. Notice if there is water condensation on cold
surfaces.
+ Keep the electrical power cord away from heated sur-
faces.
• Provide a firm, level platform for the equipment. En-
sure a stable position of the exhaust waste vapor con-
denser Peltronic
• Comply with maximum permissible pressures at in-
let and outlet and with maximum permissible pres-
sure differences between inlet and outlet. See section
„Technical data", pg. 22. Do not operate the exhaust
waste vapor condenser Peltronic
at the inlet.
• Check the power source and the condenser's rating
plate to be sure that the power source and the equip-
ment match in voltage, phase, and frequency.
• Make sure the fan's air supply is adequate, especially
if the exhaust waste vapor condenser Peltronic
stalled in an enclosure. Do not block ventilation slots.
• Connect hoses gas tight at inlet and outlet of the ex-
haust waste vapor condenser Peltronic
• Note: Flexible elements will shrink when evacuated.
page 13 of 44
. Check that all fittings are secure.
®