Shutdown & storage
Short-term:
NOTICE
Remove catchpot and drain condensate.
Important: Comply with regulations when disposing of
solvents/condensates. Recycle if possible; purify if con-
taminated.
Has the exhaust waste vapor condenser Peltronic
exposed to media which may damage the condenser's
materials or form deposits?
Check the exhaust waste vapor condenser Peltronic
(check inlet and outlet) and clean, if necessary.
Clean a contaminated exhaust waste vapor condenser
Peltronic
Long-term:
as follows:
®
- Fill the condenser with a solvent (e.g., benzene) and
allow sufficient cleaning time. Observe all regulations
concerning usage and disposal of solvents!
- Drain the solvent and dispose of in accordance with
regulations. Repeat cleaning if necessary.
- Rinse the condenser several times with alcohol in or-
der to remove all solvent residues.
- Allow the emission condenser to dry.
- Take measures as described above regarding short-
term shutdown.
- Separate the exhaust waste vapor condenser Peltronic
from the application.
- Close inlet and outlet ports (e.g., with transport caps).
- Store the exhaust waste vapor condenser Peltronic
under dry conditions.