Notes concerning the use of the exhaust waste vapor
condenser Peltronic
outlet, gas!
inlet
Use the equipment only as intended, that is, for conden-
sation of emitted vapors at the outlet of vacuum systems
and pumps (i.e. at atmospheric pressure).
During operation
➨ Adopt suitable measures to prevent the formation of
explosive or flammable mixtures, use inert gas for
venting if necessary.
➨ A dopt suitable measures to prevent the release of
dangerous, toxic, explosive, corrosive, noxious or pol-
luting fluids, vapors and gases when disposing of con-
densates.
+ Maximum permissible operation pressure:
13 - 16 psi (0.9 - 1.1 bar) (absolute).
Never block the gas outlet.
• Make sure the fan's air supply is adequate. Clean venti-
lation slots, if necessary (e.g., remove dust or clean by
suction).
• Prior to every use: Inspect the condenser for faults.
There must be no damage. Do not use damaged com-
ponents.
®
The
Peltronic
of the pumped vapors at the outlet. The con-
denser's temperature is automatically con-
trolled to +50°F (+10°C).
+ Next to 100% solvent recovery.
page 28 of 44
exhaust
waste
enables an efficient condensation
®
vapor
condenser