Upper Threshold Delay (solid-state output)
The Upper Threshold Delay controls the Activation delay: the amount of time that has to
pass uninterrupted with the probe covered to a level above the Upper Threshold Setting
before the timer expires. After the timer expires, the output signal will comply to the
setting from Menu 18 for a covered probe. Whenever the level drops below the Upper
Threshold Setting before the timer expires, the timer is restarted.
When the solid-state switch control (Menu 17) is disabled, menu 13 displays - - - -. When
the solid-state switch control is enabled, menu 13 displays the Activation delay in
seconds. As an extra identifier, an upward running d is displayed to the right of the value.
Menu
Affected
Item
by:
13
17
First select the solid-state switch output at menu 17 (contact open or contact closed).
1. Set the rotary switch to 3.
2. Press the RED (+) or BLUE (–) button to increase or decrease the value.
or: Press and hold a button to start a repeat function.
or: Press both buttons simultaneously to toggle the value between minimum (0)
and maximum (100).
Page 74
Rotary
Left
Switch
Arrow
Position
3
On
Pointek CLS500 – INSTRUCTION MANUA
Mode
Identifier
Upward
Solid-state
running d at
switch
*
right of value
control
enabled
factory setting
Solid-state
switch
control
disabled
Des-
Values
cription
Upper
Range: 0
Threshold
to 100
Delay
(seconds)
0.0
Display
shows
- - - -
7ML19985GG03