Technical Data - gbo Medizintechnik Sonostat User Manual

Table of Contents

Advertisement

Sonostat

9 Technical data

Mains voltage and
Frequency:
Power consumption:
Mains fuses:
Maximum Ultrasound
Intensity:
Accuracy of the output:
Effective radiating area
Pulse ratio:
Pulse sound frequency:
Ultrasonic frequency:
Applied part:
Type of applied part:
Protection class of probe
Mode of operation:
Protection class:
Protection class of housing:
Dimensions:
Weight:
Display:
Environmental conditions:
On request, spare parts lists and circuit diagrams can be made available for repair technical personnel.
To maintain the leakage current limits it is not allowed to repair or replace components on the PCBs.
The IEC connector is used for all-pole disconnection from the mains.
gbo Medizintechnik AG reserves the right to change design and specifications
without prior notice.
gbo Medizintechnik AG
100 - 240 V
50-60 Hz
Max 120 VA
T 2A H 250V
3W/cm ² at 1 MHz, 1.5 W / cm ² at 3.3 MHz
1 MHz: +0% / -20% all other frequencies: ± 20%
5 cm² and 2.5 cm² ± 20%
CW, 5%, 10%, 20%, 30% and 50%
100 Hz ± 5%
1 MHz ± 10% / 3.3 MHz ± 10%
Titanium surface of the ultrasonic head
BF
IP X7
Continuous operation
I
IP X0
27 cm × 19.3 cm × 28 cm (W × H × D)
4.9 kg without accessories
10.1 "TFT LCD with touch screen
Operation
Temperature range +10 ° C ... +35 ° C
of
the
relative humidity 30 ... 75%
device:
Transport
Temperature range 5 ° C ... +50 ° C
and
relative humidity <90%, non-condensing
storage:
27
Version 1.2

Advertisement

Table of Contents
loading