Monitor a Patient Out of Range of Acuity

While out of range of Acuity, the Wireless Propaq CS continues to monitor
the patient and provide local alarms or alerts at the Propaq CS as needed.
When the patient with the Wireless Propaq CS goes out of range of Acuity, do
the following:
1. A DROPOUT equipment alert occurs at the Acuity Central Station.
Acknowledge the alert at Acuity.
2. An equipment alert occurs at the Wireless Propaq CS with this
message:
NETWORK FAULT
CHECK ACUITY/NETWORK CONNECTION
Press a key on the Propaq CS to acknowledge the alert and silence the
alert tones (if tones are enabled).
When the patient with the Wireless Propaq CS returns within range of
Acuity, the Propaq CS automatically reconnects to Acuity. No clinician
intervention is required.