Wireless Connection To Acuity; Connect A New Patient - Welch Allyn Propaq CS User Manual

Hide thumbs Also See for Propaq CS:
Table of Contents

Advertisement

Wireless Connection to Acuity

Connect a New Patient

1. Make sure the Acuity network cable is not plugged into the Acuity
connector and an Acuity bedside network jack. Turn on power to the
Wireless Propaq CS. The monitor displays the powerup screen for
about 10 seconds, then displays the Main Menu. The monitor is in the
powerup patient mode with the associated settings.
2. Confirm that the Wireless Propaq CS displays the flashing message
SEARCHING in the upper left corner of the display.
Monitor is searching for
a network connection.
Within a short time the monitor displays the flashing message
CONNECTING. This indicates the monitor has associated with an access
point but is not fully connected to the Acuity network.
3. Confirm that within one minute the monitor displays these alternating
messages:
ON NETWORK and WIRELESS
This indicates the Wireless Propaq CS is connected to the Acuity
network.
4. After the network connection is established, the monitor may prompt
you to select an Acuity Unit (if your facility has more than one Acuity
unit). Scroll up or down to highlight the desired Acuity Unit, then
press
Select
5. The Wireless Propaq CS displays a list of possible patients. If your
patient has been pre-admitted to the selected Acuity unit, they will be
included in the list.
12
(Need artwork -- Propaq CS display . . .)
.