While connected to the network, the Propaq CS sends patient data to Acuity.
Acuity continuously analyzes the data and provides appropriate alarm or
alert messages at the Central Station and other network devices such as a
hallway message panel or the Propaq CS itself. Acuity also stores the patient
data for viewing or report printing.
If the Wireless Propaq CS is moved out of range or loses communication with
the FlexNet network and Acuity, it continues to monitor the patient and
display patient data. While not communicating with Acuity, the Propaq CS
continues to generate local patient alarms or alert messages. Acuity does not
perform waveform analysis or generate arrhythmia messages while the
Propaq CS is not communicating with Acuity. When the Wireless Propaq CS
is returned to within range of the FlexNet network, it automatically
reconnects to Acuity.
A Propaq CS with the Acuity hardwire connection can be identified by the
Acuity connector on the right side panel.
A Wireless Propaq CS can be identified by the blue antenna cap on the top
right corner of the monitor in front of the carrying handle. Every Wireless
Propaq CS also has the Acuity hardwire connection.
Wireless Propaq CS
If a Wireless Propaq CS monitor is communicating with the network via the
wireless interface and then the Acuity hardwire cable is used to connect the
bedside Acuity network jack and the monitor, the monitor disconnects from
the wireless connection and communicates through the hardwire connection.
There may be a short interruption in the display of patient waveforms at
Acuity while this change occurs.
Update
Propaq CS DFU 3.4X
Yellow antenna cap indicates
Wireless Propaq CS
MONITOR
3A
2AG
Right Side Panel
DEFIB SYNCHRO
!
!
EKG x 1000
Acuity
!
hardwire
!
connector
12-28V, 3A
7