Hardwired Connection to Acuity

Connect the Monitor to the Acuity System

1. If the Propaq CS monitor has already been connected to the patient,
save the patient's Trends and Alarm Limit settings by keeping the
monitor turned on.
The monitor transmits up to five hours of trend information when you
connect it to the Acuity network.
If the monitor has not been connected to the patient, clear any prior
patient's trends and alarm limit settings by turning off the monitor,
then turning it on after a few seconds.
2. If the monitor is not already connected to the patient, attach leads and
sensors to the patient as described in the Propaq CS Directions For Use.
3. Plug in the Acuity network cable to the
Acuity network jack on the monitor side
panel as shown. Plug in the other end of the
cable to the bedside Acuity network jack.
4. Connect the ac power adapter to the
monitor and the wall outlet to charge the
battery. Check to see that the green battery
charging light on the monitor's right side
panel is on.
5. When the monitor completes the
connection to Acuity, it displays the
message ON NETWORK (or alternating
messages ON NETWORK and SERIAL if
the monitor has wireless capability).
Confirm the patient identification at the
bedside or enter the patient information at the Acuity Central Station
using the Patient ID Setup Window.
6. If alarm limits have not been set, do so at the monitor or at the Acuity
Central Station using the Alarms Setup Window.
Update
Propaq CS DFU 3.4X
MONITOR
DEFIB SYNCHRO
!
!
3A
2AG
!
12-28V, 3A
Acuity Connector
Battery charging light
Power input connector
EKG x 1000
!
9