C
LEANING THE
P
ATIENTS
The mask system (excluding the valve
membrane) is validated for multipatient
use. Clinicians should refer to the Masks
Disinfection/Sterilization Guide for details.
To clean the headgear between patients
use the method described in the previous
section "Cleaning the Mask in the Home".
R
EPROCESSING THE
P
BETWEEN
ATIENTS
The mask should be reprocessed when
used between patients.
Cleaning, disinfection and sterilisation
instructions are available from the ResMed
website, www.resmed.com/masks/
sterilization.
If you do not have Internet access, please
contact your ResMed representative.
CAUTION
The valve membrane must be replaced
with a new valve membrane when the
mask is reused on another patient. This
part cannot be disinfected or sterilized.
R
EASSEMBLING THE
ATTENTION
It is important to follow the instructions
detailed in this section. Any deviation
from the specified order of assembly
could affect the functioning of your mask.
Reassembling the entire mask involves two
stages:
1. Valve Reassembly
2. Mask Reassembly
V
R
ALVE
EASSEMBLY
See section E on the illustrations sheet.
M
ASK BETWEEN
M
ASK
M
ASK
Note: This stage must be completed before
Mask Reassembly.
Make sure all parts of the mask system,
especially the valve membrane, are clean
and dry before reassembling. If the valve
membrane is damaged, distorted or torn,
replace it with a new valve membrane.
1 Insert the valve membrane into the
valve frame and gently push it into place 
(Fig. E-1). Make sure it is seated evenly
and flatly around the rim of the valve
frame, as the membrane may produce a
noise while you are breathing out, if it is
fitted incorrectly.
2 Push the valve frame over the tabs of
the valve elbow until they lock together 
(Fig. E-2). Make sure the valve frame and
valve elbow are securely connected to
each other.
3 Insert the swivel into the end of the
valve elbow and secure it with the swivel
clip (Fig. E-3).
4 Set the assembled valve (Fig. E-4) aside
for use in the following stage.
M
R
ASK
EASSEMBLY
See section F on the illustrations sheet.
1 Check that the holes in the exhaust vent
are not obstructed. If they are
obstructed, replace the exhaust vent.
Push the exhaust vent onto the mask
frame so the small round bump on the
exhaust vent is on the outside of the
mask frame (Fig. F-1).
CAUTION
The exhaust vent must be correctly
inserted into the mask frame to
prevent leakage. Poor sealing may
reduce pressure and affect therapy.
2 Twist the luer lock port caps onto the
mask ports (Fig. F-2).