Disassembly and repair

These procedures provide instructions for monitor disassembly and board removal. Except where
otherwise noted, the assembly procedure is the reverse of the disassembly procedure.
An exploded view of the assembly precedes disassembly instructions with callouts referencing the
parts. In the instructions, numbers in parenthesis refer to callouts in the exploded drawings.
Each part's disassembly instructions may include one or both of the following:
Reassembly notes: This contains information specific to reassembly not addressed in the
disassembly instructions.
When replacing the component: This contains information specific to installing a new
option or replacement part.
For information about screws or connectors used in the monitor, see "Screws" and "Connectors"
in the appendices.
Note
After performing any of these procedures and prior to returning the monitor to
service, you must use the service tool, Gold edition, to complete the full suite of
functional tests to ensure that all systems are operating within the design
specifications. For more information about these tests and the service tool, see
"Functional verification and calibration."
If you do not have the service tool, contact Welch Allyn Technical Support.
WARNING Electrical shock hazard. Disconnect AC power before opening the
monitor. Disconnect and remove the battery before proceeding with disassembly.
Failure to do this can cause serious personal injury and damage to the monitor.
WARNING Risk of fire, explosion and burns. Do not short-circuit, crush,
incinerate, or disassemble the battery pack.
WARNING Safety risk. Do not attempt to service the monitor when the monitor is
connected to a patient.
Caution Before disassembling the monitor, disconnect the AC power cord and
any attached accessories (SpO2 sensors, blood pressure hoses and cuffs,
temperature probes, and accessories) before disassembly.
Caution Remove the probe well in the temperature module prior to disassembly.
Caution Perform all repair procedures at a static-protected station.
Caution When the monitor case is opened, regard all parts as extremely fragile.
Execute all procedure steps with care and precision.