Reference Channel - Ametek 7280 Instruction Manual

Wide bandwidth dsp lock-in amplifier
Table of Contents

Advertisement

Chapter 6, COMPUTER OPERATION
AUTOMATIC [n] AC Gain automatic control
LF [n

6.4.02 Reference Channel

REFMODE [n]
6-12
n
AC Gain
0
0 dB
1
6 dB
2
14 dB
3
20 dB
4
26 dB
5
34 dB
6
40 dB
7
46 dB
8
54 dB
9
60 dB
10 66 dB
The value of n sets the status of the AC Gain control according to the following table:
n
Status
0
AC Gain is under manual control, either using the front panel or the ACGAIN
command
1
Automatic AC Gain control is activated, with the gain being adjusted according
to the full-scale sensitivity setting
n
]
Signal channel line frequency rejection filter control
1
2
The LF command sets the mode and frequency of the line frequency rejection (notch)
filter according to the following tables:
n
Selection
1
0
Off
1
Enable 50 or 60 Hz notch filter
2
Enable 100 or 120 Hz notch filter
3
Enable both filters
n
Notch Filter Center Frequencies
2
0
60 Hz (and/or 120 Hz)
1
50 Hz (and/or 100 Hz)
Reference mode selector
The value of n sets the reference mode of the instrument according to the following
table:
n
Mode
0
Single Reference / Virtual Reference mode
1
Dual Harmonic mode
2
Dual Reference mode
NOTE: When in either of the dual reference modes the command set changes to
accommodate the additional controls. These changes are detailed in section 6.4.14