Signal Channel - Ametek 7280 Instruction Manual

Wide bandwidth dsp lock-in amplifier
Table of Contents

Advertisement

Chapter 6, COMPUTER OPERATION

6.4.01 Signal Channel

IMODE [n]
VMODE [n]
CP [n]
FLOAT [n]
SEN [n]
SEN.
6-10
Current/Voltage mode input selector
The value of n sets the input mode according to the following table:
n
Input mode
0
Current mode off - voltage mode input enabled
1
Normal current mode enabled - connect signal to B input connector
2
Low noise current mode enabled - connect signal to B input connector
3
High bandwidth current mode enabled - connect signal to B input connector
If n = 0 then the input configuration is determined by the VMODE command.
If n > 0 then current mode is enabled irrespective of the VMODE setting.
Voltage input configuration
The value of n sets up the input configuration according to the following table:
n
Input configuration
0
Both inputs grounded (test mode)
1
A input only
2
-B input only
3
A-B differential mode
Note that the IMODE command takes precedence over the VMODE command.
Input connector coupling mode control
The value of n sets up the input coupling mode according to the following table:
n
Coupling mode
0
Fast
1
Slow
Input connector shield float/ground control
The value of n sets the input connector shield switch according to the following table:
n
Selection
0
Ground
1
Float (connected to ground via a 1 kΩ resistor)
Full-scale sensitivity control
The value of n sets the full-scale sensitivity according to the following table,
depending on the setting of the IMODE control: