- If loss of electrical power during use is likely to
generate an unacceptable risk, the device must be
connected to a suitable power source (e.g. UPS).
Using the device:
- Do not use the device if it appears to be damaged
or faulty.
- Turn the device off before unplugging the power
cord.
- To unplug the power cord, grip the cord plug and
hold the wall socket.
- Never use any other irrigation solution containers
than those intended for suspension from the
supplied brackets.
- The device must only be used with bottles or bags
of physiological saline or sterile water.
- The capacity of the irrigation solution containers
used must not exceed one litre.
- When the device is not to be used for a long period
of time, unplug the device from the electric supply.
- Do not exert excessive force on the screen.
- Do not move I
MPLANT
Environment:
- Do not cover the device or obstruct the ventilation
vents.
- Do not immerse the device in liquid, and do not use
it outdoors.
- Do not tilt the device at an angle greater than 5°.
- Do not place the device near a heat source.
- Make sure that the cords are not in a traffic path.
- The device must be stored in its original packaging
in an appropriate and safe place.
- The device is not designed for use in the presence
of anesthetic gases or any other flammable gas.
- Do not expose the device to water vapor, or
splashes.
- Any condensation inside an electrical device is
potentially dangerous.
- If the device is to be moved from a cold
environment to a hot one, it must not be used until
it has reached room temperature.
- The device is not designed to work near ionizing
radiation.
- Do not insert metal objects into the device (risk of
6
C
2 during use.
ENTER
electric shock, short-circuit or projection of
hazardous substances).
Maintenance:
- Before and after each use, your I
must be disinfected with SATELEC-approved agents.
- Before each procedure, it is essential to make sure
that the accessories to be used have been cleaned,
disinfected and sterilized.
Accessories:
- The device has been designed and developed for
use only with SATELEC accessories to ensure
maximum safety and performance.
- Use of accessories from other manufacturers is a
potential hazard for you or your patients.
Repair:
- Warning: Do not repair or modify the device without
prior authorization from SATELEC.
- Warning: If the device is modified or repaired,
specific checks and tests must be performed to
ensure that the device can still be used safely.
- In the case of a fault, contact the supplier of your
device. Do not use unauthorized repairers, who
might make your device dangerous for you and your
patients.
If in any doubt, contact an approved dealer or our
customer support department:
- www.acteongroup.com
- E-mail: satelec@acteongroup.com.
III - DESCRIPTION

3. 1 PHYSICAL DESCRIPTION

The I
C
2 has the following components:
MPLANT
ENTER
- a control unit (Fig. 1),
- a multifunction footswitch (Fig. 11),
- a micro-motor cord with connectors (Fig. 1-2),
- two brackets for irrigation solution (Fig. 1-3),
- an I-S
L
micro-motor (Fig. 1-4) without
URGE
ED
contra-angle,
C
2
MPLANT
ENTER

Advertisement