Fault
Aucun fonctionnement
(Ecran LCD éteint)
Device does not operate
(LCD screen off)
Device does not operate
(LCD screen on)
Motor does not operate
Torque fault
Speed fault
No spray
Possible cause
Power cord not plugged in properly.
Power switch in position O (OFF).
No power voltage.
Power receptacle fuse(s) blown.
Internal fuse blown
Transmission fault.
Motor or motor connections problem.
Thermal cut-out.
Incorrect torque adjustment.
Wrong contra-angle.
Incorrect speed adjustment.
Wrong contra-angle.
Irrigation solution bag or bottle empty.
Irrigation off.
Irrigation line clogged.
Line end piece on contra-angle clogged.
Remedy
Check the electric socket.
Return to SATELEC after-sales.
Place the power switch in position I (ON).
Call an electrician.
Replace the power receptacle fuses.
Return to SATELEC after-sales.
Switch the device OFF then ON again.
Return to SATELEC after-sales.
Check the motor cord connector.
Check that the cord connector is properly
plugged into the motor connector of the
device.
Return to SATELEC after-sales.
Allow the micro-motor to cool down.
Adjust the torque in accordance with good
dental practice.
Change the contra-angle.
Return to SATELEC after-sales.
Adjust the speed in accordance with good
dental practice.
Change the contra-angle.
Return to SATELEC after-sales.
Replace the bag or bottle of irrigation
solution.
Press the irrigation ON/OFF button.
Change irrigation line.
Clear the contra-angle line end piece.
23