In the [Diagnostic Data] Menu, highlight [Path Select] and select the desired transducer path.
Press <ENT> to select path.
Press the <Down Arrow> and scroll to the [Site Setup Data] menu cell. Press the <Right
Arrow> to select it.

"Manual" Adjustment Procedure

In the [Site Setup Data] Menu, press the <Down Arrow> and scroll to the [HF] menu cell. Press
the <Right Arrow> and a pop-up [Manual] prompt will appear as shown below. Note: Press the
<Up/Down Arrow> to select [Automatic], if desired.
Use the numerical keys to input the desired correction value. Press <ENT> to input value.
Siemens
Dual Path
fx (drive)
N (burst length)
Ltn in
Vf max CU
Vs max m/s
Vs min m/s
Samples/Cycle
Max Damping
Min Damping
HF
Site Setup Data
The new correction value will appear next to the [HF] menu cell as shown below.
Dual Path
4SS10G
4SS10G
29
5
5.316
Adjustment
>Manual
1355.00
29
>-0.000
Siemens
Transducer Setup Data
Path Select
Path Enable
Flow Rate
Application Info
Liquid Data
Site Seup Data
Test Facilities
Print Site Setup
Date Site Created:
Diagnostic Data
Siemens
fx (drive)
N (burst length)
Ltn in
Vf max CU
Vs max m/s
Vs min m/s
Samples/Cycle
Max Damping
Min Damping
HF
Site Setup Data
2-64
1010PFM-3G
Dual Path
4SS10G
1
Yes
No
04.09.0314.52.23
Dual Path
4SS10G
29
5
5.316
>Manual
= -0.120
1355.00
29
>-0.000