Vision screener. troubleshooting malfunctions and measuring interruptions
Hide thumbs Also See for S12C:

Advertisement

h) Corneal reflexes are too dark!
In this example, too little infrared light is reflected from the retina.
Cause:
This error message is displayed if the corneal reflexes are too dark.
Reason:
The measurement is carried out with infrared light which is reflected from the
retina. If too little light comes back from the retina, no measurement can be
taken.
Tip:
Darken the examination room so that the pupils become larger and more infrared
light can return to the camera.