5.9.3 Installation and connection of control button and power supply

Before connection of the control button and power supply, it is necessary to dismount the covers on the motor
box.
The covers are dismounted by loosening the two screws, which holds each cover and pulling the covers
horizontally to the side.
The control switch is mounted in the room and at the height, the user prefers.
The control switch is opened with tools, and the back side is mounted on the wall using two screws Ø3.5X30mm.
The wire is shortened to a suitable length and connected to the motor box. See drawing pos 1.
The power supply is mounted on the wall, visible or hidden, depending on what is most suitable.
The power supply is mounted on a Ø3.5X30mm screw on the wall, using the "key" hole on the back side of the
power supply.
At the bottom of the power supply there is hole for a Ø3.5X30mm screw which makes it possible to lock
the power supply on the wall.
The wire is shortened to a suitable length and connected to the motor box. See drawing pos 2.
The cover is re-attached, when the wires are connected correctly and the control switch is tested.
Please remember to mount the wires in the end-plate, in order to avoid that the wire conflicts with the cover
when it is re-attached. See drawing pos 3.
1
2
3
NB! It is very important that there are no loose wires hanging from the coupling.