Cleaning Components - Alcatel I series User Manual

Rotary vane pumps
Table of Contents

Advertisement

Cleaning components
Cleaning metal componeonts
Solvents are required to clean components.
Standard precautions should be taken in compliance with the manufacturer's
instructions.
After use in mineral or synthetic oil,
clean the metal components with a
mineral products based solvent such as
AXARELil) , CARECLEANI2),
PREMACLEANI3), NAPHTEOLl4).
Proceed as follows:
• Clean when cold or hot (max.
45°q
by dipping or using a cloth
• Vacuum dry in a ventilated oven
• The component must be cleaned a second time with alcohol.
After use in (perfluorinate) synthetic oil,
clean the metal components in a
solvent such as
GALDEN S
90™15)
and proceed as follows:
• Clean when cold by dipping or using a cloth
• Dry the components in the air or with compressed air
After use in (non-perfluorinate) synthetic or mineral oil,
clean the metal
components with a solvent such as alcohol and proceed as follows:
• Clean when cold by dipping or using a cloth
• Dry the components in the air
• )ndustrial washing solutions can also be used. The cleaning operation
should be followed by vacuum drying.
Cleaning the oil level sight
glass
94
I, SO, C2 series pumps
When cleaning this plastic sight glass, avoid contact with alcohol or alcohol-
based washing solutions. Clean the component with a solvent, but do not
steep it, and rinse it immediately.
C1 series pumps
The sight glass of these pumps is made of glass.
Ii!
13)
14)
15)
DUPONT DE NEMOURS registered trademark
CASTROl registered trademark
DOW registered trademark
Nippon Chemical registered trademark
MONTEDISON registered trademark
, ,