AC Voltage Gain Adjustment Procedure

Review the
"Test Considerations"
Considerations"
on page 90 sections before beginning this procedure.
ACV
1 Press
to enter AC Voltage Gain Calibration.
2 The primary display will show the uncalibrated value and the
secondary display will show the reference value of Cal Item.
3 Configure each Cal Item shown in the adjustment table below.
4 Use
(Auto) or
5 Apply the input signal shown in the Input and Frequency columns of
the table below.
Always complete tests in the same order as shown in the appropriate table.
6 Enter the actual applied input (see
page 86).
Disp
to start the adjustment. The CAL annunciator flashes to
7 Press
indicate the calibration is in progress.
Successful completion of each adjustment value is indicated by a
short beep and the primary display briefly showing PASS.
An adjustment failure is indicated by a long beep, the primary
display showing FAiL and a calibration error number appearing in
the secondary display. Check the input value, range, function, and
entered adjustment value to correct the problem and repeat the
adjustment step.
8 Repeat steps 3 through 7 for each gain adjustment point shown in the
table.
9 Verify the AC Voltage Gain adjustments using the