Ac Voltage Gain Adjustment Procedure - Agilent Technologies 34405A User's And Service Manual

5 1/2 digit multimeter
Hide thumbs Also See for 34405A:
Table of Contents

Advertisement

4
Performance Tests and Calibration
Table 16
DC Current Gain Adjustment

AC Voltage Gain Adjustment Procedure

N O TE
100
Input
Caution: Connect calibrator to multimeter's 12A and LO terminals before applying 10A
10 A
Review the
"Test Considerations"
Adjustment Considerations"
beginning this procedure.
A C V
1 Press
2 The primary display will show the uncalibrated value and
the secondary display will show the reference value of Cal
Item.
3 Configure each Cal Item shown in the adjustment table
below.
4 Use
Item.
5 Apply the input signal shown in the Input and Frequency
columns of the table below.
Always complete tests in the same order as shown in the appropriate
table.
6 Enter the actual applied input (see
Values"
on page 91).
on page 71 and
on page 95 sections before
to enter AC Voltage Gain Calibration.
(Auto) or
34405A User's and Service Guide
Function
Cal Item
10 A
"Gain
(Range) to select the Cal
"Entering Adjustment