Always complete tests in the same order as shown in the appropriate table.
6 Enter the actual applied input (see
page 86).
Disp
to start the adjustment. The CAL annunciator flashes to
7 Press
indicate the calibration is in progress.
Successful completion of each adjustment value is indicated by a
short beep and the primary display briefly showing PASS.
An adjustment failure is indicated by a long beep, the primary
display showing FAiL and a calibration error number appearing in
the secondary display. Check the input value, range, function, and
entered adjustment value to correct the problem and repeat the
adjustment step.
8 Repeat steps 3 through 7 for each gain adjustment point shown in the
table.
9 Verify the DC Current Gain adjustments using the
Verification Test"
Table 16
DC Current Gain Adjustment
Input
Remove test leads from Input terminals
10 mA
100 mA
1000 mA
Caution: Connect calibrator to multimeter's 12A and LO terminals before applying 10A
10 A
manuals search engine
Performance Tests and Calibration
"Entering Adjustment Values"
on page 74.
"DC Current Gain
Function
Cal Item
DC Current
Open
10 mA
100 mA
1000 mA
10 A
4
on
93