To Replace the Power Line Fuse

34405A User's and Service Guide
Downloaded from
www.Manualslib.com
The power line fuse is located within the multimeter's fuse–holder
assembly on the rear panel. The multimeter is shipped from the factory
with a power–line fuse installed. The supplied fuse is a time- lag,
low- breaking, 0.2A/ 250V, 5x20mm fuse, Agilent part number 2110- 1395. If
you determine that the fuse is faulty, replace it with one of the same size
and rating.
1 Disconnect power cord. Depress tabs 1 and
2 and pull fuse holder from rear panel.
3 Rotate line voltage selector and reinstall so
correct voltage appears in fuse holder
window.
100, 120, 220 (230) or 240 Vac
manuals search engine
Disassembly and Repair
2 Remove the line voltage selector from fuse
holder assembly.
2
1
4 Replace fuse holder assembly in rear
panel.