Performance Tests and Calibration
C A U T I O N
90
Downloaded from
www.Manualslib.com
Gain Adjustment Considerations
• The zero adjustment procedure must have been recently performed
prior to beginning any gain adjustment procedures.
• Be sure to allow the instrument to warm up and stabilize for 2 hours
before performing the adjustments.
• Consider the thermal effects as you are connecting test leads to the
calibrator and multimeter. It is recommended to wait one minute before
starting the calibration after connecting the test leads.
Never turn off the instrument during a Gain Adjustment. This may cause calibration
memory for the present function to be lost.
Valid Gain Adjustment Input Values
using the following input values.
Table 14
Valid Gain Adjustment Input Values
Function
Range
DC Volts
100 mV, 1 V, 10 V, 100 V, 1000 V
DC Current
10 mA, 100 mA, 1000 mA, 10 A
100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1MΩ, 10 M
Ohms
Ω
Frequency
Autorange/1 kHz
AC Current
1 mA, 10 mA, 100 mA, 1000 mA, 10 A
AC Volts
10 mV, 100 mV, 1 V, 10 V, 100 V, 750 V
Capacitance
0.4 nF, 1 nF, 10 nF, 100 nF, 1 µF, 10 µF,
100 µF, 1000 µF, 10000 µF
manuals search engine
Gain adjustment can be accomplished
Valid Amplitude Input Values
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
Input
100 mV rms, 900 Hz to 1100 Hz
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
34405A User's and Service Guide