Agilent Technologies 34405A User's And Service Manual page 96

5 1/2 digit multimeter
Hide thumbs Also See for 34405A:
Table of Contents

Advertisement

4
Performance Tests and Calibration
96
Downloaded from
www.Manualslib.com
6 Enter the actual applied input (see
page 86).
Disp
to start the adjustment. The CAL annunciator flashes to
7 Press
indicate the calibration is in progress.
Successful completion of each adjustment value is indicated by a
short beep and the primary display briefly showing PASS.
An adjustment failure is indicated by a long beep, the primary
display showing FAiL and a calibration error number appearing in
the secondary display. Check the input value, range, function, and
entered adjustment value to correct the problem and repeat the
adjustment step.
8 Repeat steps 3 through 7 for each gain adjustment point shown in the
table.
9 Verify the AC Current Gain adjustments using the
Verification Test"
Table 18
AC Current Gain Adjustment
Input
Current
Frequency Function
1 mA
1kHz
AC Current 1 mA
10 mA
1kHz
100 mA
1kHZ
1000 mA 1kHz
Caution: Connect calibrator to multimeter's 12A and LO
terminals before applying the following 1A and 10A
1 A
1kHz
10 A
1kHz
manuals search engine
"Entering Adjustment Values"
on page 78.
Frequency as 1kHz
Cal Item
10 mA
100 mA
1000 mA
1 A
10 A
"AC Current
34405A User's and Service Guide