Table 9
AC Volts Verification Test
Vrms
Input Frequency
100mV
100mV
100mV
1V
1V
1V
10V
10V
10V
10V
100V
100V
100V
750V
Caution: Set the calibrator output to 0V before disconnecting it from the multimeter input terminals.
34405A User's and Service Guide
Downloaded from
www.Manualslib.com
AC Voltage Verification Test
Configuration: AC Volts (CONFigure[:VOLTage]:AC)
1 Select the AC Voltage function.
2 Select each range in the order shown below. Provide the indicated input
voltage and frequency. Compare measurement results to the appropriate
test limits shown in the table. (Be certain to allow for appropriate
source settling.)
Function
1kHz
AC Voltage
30kHz
100kHz
1kHz
30kHz
100kHz
45Hz
1kHz
30kHz
100kHz
1kHz
30kHz
100kHz
1kHz
[1] Select 5½- digit measurement resolution
Q = Quick performance verification test points
manuals search engine
[1]
Range
Quick Check
100mV
100mV
100mV
1V
1V
1V
10V
10V
10V
10V
100V
100V
100V
750V
Performance Tests and Calibration
Error from Nominal
1 year
± 0.3 mV
± 1.8 mV
± 5.3 mV
Q
± 3.0 mV
± 11 mV
± 32 mV
± 110 mV
± 30 mV
Q
± 0.11 V
± 0.32 V
Q
± 0.3 V
± 1.1 V
± 3.2 V
± 2.25 V
4
77