Gain Verification

This test checks the full- scale reading accuracy of the instrument.
Verification checks are performed only for those functions and ranges with
unique gain calibration constants.
DC Voltage Gain Verification Test
1 Connect the calibrator to the front panel HI and LO input terminals.
2 Select each function and range in the order shown below. Provide the
input shown in the table below.
3 Make a measurement and observe the result. Compare measurement
results to the appropriate test limits shown in the table. (Be certain to
allow for appropriate source settling when using the Fluke 5520A.)
[1]
Function
Range
DC Volts
100mV
100mV
1 V
1 V
10 V
100 V
1000 V
[1] Select 5½- digit measurement resolution
Q = Quick performance verification test points
manuals search engine
Performance Tests and Calibration
Error from Nominal
Quick Check
1 year
±33 µV
±33 µV
Q
±0.31 mV
±0.31 mV
±3.0 mV
Q
±30 mV
±0.3 V