Gain Verification; Table 2-3 Dc Voltage Gain Verification - Agilent Technologies U2741A USB Service Manual

Modular 5.5 digits
Hide thumbs Also See for U2741A USB:
Table of Contents

Advertisement

2
Performance Test Calibration

Gain verification

Table 2-3 DC voltage gain verification

Input
100 mV
–100 mV
1 V
–1 V
10 V
100 V
300 V
18
This test checks the full- scale reading accuracy of the
instrument. Verification checks are performed only for those
functions and ranges with unique gain calibration constants.
DC voltage gain verification test
DC voltage configuration:
(CONFigure[:VOLTage][:DC] [<range|MIN|MAX|AUTO>
[,resolution|MIN|MAX|DEF>]])
1 Connect the calibrator to the front panel HI and LO input
terminals.
2 Select each function and range based on the input voltage
as shown in
3 Make a measurement and observe the result. Compare the
measurement results with the appropriate test limits as
shown in
source settling when using the Fluke 5520A.)
[1]
Function
Range
100 mV
100 mV
1 V
DC Voltage
1 V
10 V
100 V
300 V
[1]
Select NPLC 20 digit resolution.
Q = Quick performance verification test points
U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide
Table 2-
3.
Table 2-
3. (Be certain to allow for appropriate
Quick check
Q
Q
Error from nominal
1 year
±23 µV
±23 µV
±0.2 mV
±0.2 mV
±2.3 mV
±23 mV
±69 mV