Canceling an exposure

5.1.3.5
An exposure that has been triggered can be canceled again at any time.
CAUTION
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it to
the starting position.
The exposure is released.
1. Let go of the exposure release button.
The exposure is immediately terminated.
Message H320 appears.
The exposure data of the terminated exposure is displayed in
Sidexis.
The radiation time and dose-area product flash.
2. Guide the patient out of the unit.
3. Press the R key.
The actual required radiation time is acknowledged.
4. Press the R key again.
The rotating unit moves to the starting position.
The unit is now ready for the next exposure.
IMPORTANT
The program settings must be checked before the exposure is repeated.
Any changed program settings must be preselected again.
5 Operation
5.1 Acquiring the X-ray image