NextSeq 550Dx Research Mode Instrument Reference Guide
RTA Errors
To troubleshoot RTA errors, first check the RTA error log, which is stored in the RTALogs folder. This
file is not present for successful runs. Include the error log when reporting issues to Illumina Technical
Support.
Troubleshooting Files for Array Scans
Key File
Scan parameters file
(ScanParameters.xml)
Log files
Metrics files
Rescan file

Resolve Automatic Check Errors

If errors occur during the automatic check, use the following recommended actions to resolve the error.
Automatic checks differ for sequencing and array scans.
Checks for Sequencing Runs
If a pre-run check fails, the reagent cartridge RFID is not locked and can be used for a subsequent run.
However, the RFID is locked after the foil seals have been pierced.
System Checks
Recommended Action
Doors Closed
Make sure that the compartment doors are closed.
Consumables
Consumable sensors do not register. Make sure that each consumable is properly loaded.
On the run setup screens, select Back to return to the loading step, and repeat run setup.
Loaded
Required Software
Critical components of the software are missing.
Perform a manual software update to restore all software components.
Instrument Disk
The instrument hard drive does not have sufficient disk space to perform a run. It is possible
Space
that data from a previous run did not transfer.
Clear run data from the instrument hard drive.
Network Connection
The network connection has been interrupted. Check network status and the physical network
connection.
Network Disk Space
Either the BaseSpace account is full or the network server is full.
Temperature
Recommended Action
Temperature
Contact Illumina Technical Support.
Temperature
Contact Illumina Technical Support.
Sensors
Fans
Contact Illumina Technical Support.
Document # 1000000041922 v01
For Research Use Only. Not for use in diagnostic procedures.
Folder
Description
Root
Contains information about scan parameters. Information includes scan date,
folder
BeadChip barcode, cluster file location, and manifest file location.
Logs
Log files describe each step performed on the instrument during the scan.
[Barcode]
Metrics are provided as sample metrics and as section metrics.
[barcode]_sample_metrics.csv—For each sample and channel (red and green), lists
Percent Off Image, Percent Outliers, P05, P50, P95, Avg FWHM Avg, FWHM Stddev,
and Min Registration Score.
[barcode]_section_metrics.csv—For each section and tile, lists Laser Z-position,
Through Focus Z-position, Red FWHM, Green FWHM, Red Avg Pixel Intensity, Green
Avg Pixel Intensity, Red Registration Score, and Green Registration Score.
[Barcode]
[barcode]_rescan.flowcell—Lists the tile locations adjusted for a rescan, which include
an increased tile-to-tile overlap.