illumina NextSeq 550Dx Instrument Reference Manual page 46

Research mode
Table of Contents

Advertisement

NextSeq 550Dx Research Mode Instrument Reference Guide
Restart to Diagnostic Mode
Use the Restart command to safely shut down the instrument and reboot to diagnostic mode.
Diagnostic mode is the default boot mode.
1
Select Manage Instrument.
2
Select Shutdown Options.
3
Select Restart.
Shut Down the Instrument
1
Select Manage Instrument.
2
Select Shutdown Options.
3
Select Shut Down.
The Shut Down command safely shuts down the software and turns off instrument power. Wait at
least 60 seconds before turning on the instrument again.
NOTE
By default, the instrument boots in diagnostic mode when turned on.
CAUTION
Do not relocate the instrument. Moving the instrument improperly can affect the optical alignment and
compromise data integrity. If you have to relocate the instrument, contact your Illumina representative.
Exit to Windows
The Exit to Windows command provides access to the instrument operating system and any folder
on the instrument computer. The command safely shuts down the software and exits to Windows,
1
Select Manage Instrument.
2
Select Shutdown Options.
3
Select Exit to Windows.
Document # 1000000041922 v01
For Research Use Only. Not for use in diagnostic procedures.
41