8 • MAINTENANCE
After 2,000 hours of operation, KINETEC Prima Advance requires a few lubrication and maintenance operations (lubrication of the
joints, pointer stops and ball screws). When the machine is switched on the yellow light blinks 5 times to indicate that the service
interval has been reached.
Despite that warning, you can continue to use your KINETEC Prima Advance by pressing START, but you should contact your nearest
KINETEC technician to have the maintenance operations conducted as soon as possible.
9 • CLEANING
Before conducting any cleaning operation, SWITCH the unit OFF and disconnect the power supply.
In order to ensure optimal hygiene, you are advised to clean the machine for each new patient.
Use a DISINFECTANT (PROPANOL/ISOPROPANOL or ALDEHYDE-based solution). Spray the disinfectant on the SURFACES
(plastic shells and metal components).
To clean the KINETEC hygienic pads, please refer to paragraph 5.
10 • ELIMINATION AND RECYCLING
a • Packing: Packing must be separated from the components plastic and paper/cardboard and given to the specific
sites from recycling.
b • KINETEC PATIENT PAD KIT: To clean with a product of disinfection then to give it to the specific sites of recycling.
c • Prima Advance unit: It contains electronic components, cables, aluminum, steel and plastic parts. When the splint is
not operational any more, to dismount and separate in groups from materials and to give them to correct unit of
recycling or to turn over the machine to Kinetec for destruction.
11 • TROUBLESHOOTING
A spare parts list and technical catalog are available to you on request from your KINETEC distributor.
If, after connecting the power supply cable to the power supply and switching on KINETEC Prima Advance:
The light is off:
- Check that the electrical socket is live using another device.
- Replace the fuse(s) of the connector (2) with fuses of the same type and caliber: 2 fuses T 500 mA 250V (6.3 x 32).
- If the led still does off, contact your nearest KINETEC technician.
If the machine fails to operate but the yellow light remains on,
- Press the START/STOP switch 1 more times.
Your KINETEC Prima Advance still does not function: Contact your nearest KINETEC technician.
Your KINETEC Prima Advance does not function and the led indicates:
The LED blinks 1 time
The LED blinks 2 times
The LED blinks 3 times
The LED blinks 4 times
The LED blinks 5 times
The LED blinks 6 times
The LED blinks 7 times
Contact your nearest KINETEC technician.
12 • TECHNICAL SPECIFICATIONS
Product
Weight: 11,8 Kg
Splint dimensions:
94 cm (37 in) x 33 cm (13 in) x 33cm (13
in)
Angular limits: -5 to 115°
Speed: from 40 to 145° per minute.
Patient height: from 1.5 m (4,92 ft) to 1.95
m (6,40 ft)
13 • SYMBOLS USED
Caution
(consult the accompanying
documents
TYPE B device
(protection against electric
shocks)
STOP (power off)
ON (power on)
Prima Advance
angle measurement function failure
no movement
abnormal consumption
the motor PCB provides the energy but motor consumption is null.
Service Time ≥ 2000h
not enough power
START/STOP switch failure.
Electricity
Power supply: 100-240 V~
Frequency: 50-60Hz
Power consumption: 50VA
Device of type B class I
IP 20.
PCB is powered (12Vdc)
Service
Minimum SPEED
Maximum SPEED
Environment
- Storage/transport conditions:
Temperature: -40 to 70°C / -40 to 160°F
Relative humidity: up to 90%
- Operating conditions:
Room temperature: 10 to 40°C/50 to 105°F
Relative humidity: up to 80%
GB
FLEXION limit
EXTENSION limit
Contains electric and
electronic components; not to
throw in the dustbins of
household refuse
2