EI/DI display field
Exposure menu list
(one-image display)
/
2010. 10 897N0519K
: Displays a menu for which the patient information is modified.
: Adds exposure menus and displays the "Exposure Menu Addition box".
: Changes exposure menus and displays the "Exposure Menu Change box".
: Performs auto-repeat exposure. Exposures can be performed repeatedly using the
exposure menu same as the selected one.
: Handles a read image as a rejected one.
: Adds the same exposure menu as the selected one.
: Deletes exposure menus.
: Scrolls the exposure menu list in one-image display mode and the image display field
in six-image display mode.
: Before exposure (image reading session), checks and changes, as necessary, exposure
parameters used for the exposure menu. The exposure parameter selection box is
displayed.
: Changes film format. The format change is possible either before or after image
reading.
: Displays images in monitor mode.
: Switches display mode alternately to six-image display mode.
NOTE
In six-image display mode, only up to 18 exposure menus can be displayed.
: Edits image information.
: When an image is displayed (or is selected in six-image
display mode), the EI and DI (defined in IEC 62494-1: 2008)
are displayed.
The EI and DI can be displayed only for normal images
(except mammography images). They do not appear for
energy subtraction images, long-view images, etc. When an
image is replaced with other one, the values are updated
accordingly.
The EI/DI display field can be repositioned by moving the □
mark.
Whether the EI/DI display field is displayed or not depends
on the setting made at the time of installation.
NOTES
• In case that a cassette-type Image Reader is used, the EI
and DI values are displayed properly only when the type
of the IP is ST-VN, ST-VI, or ST-BD.
• When using the EI and DI values for image quality
management, calibrations of Image Readers should be
carried out periodically.
• The EI and DI may vary if a wrong exposure menu was
selected.
: This list shows exposure menus selected on the
"Exposure Menu Selection Screen". Moving up and down
the "Exposure menu list" with any exposure menus selected
accordingly will allow you to change order of selected
exposure menus.