Alarm Conditions - CAIRE FreeStyle Comfort Technical Manual

Hide thumbs Also See for FreeStyle Comfort:
Table of Contents

Advertisement

Alarm Conditions

Priority
Alarm
Description
1
System
Malfunction
Battery over
Temp
2
System
Malfunction
High
Temperature
3
System
Malfunction
High
Pressure
4
No Breath
Detected
Display Screenshot
FreeStyle Comfort
Display
Audible
Category
Signal
Alarm
1 beep
every 20
sec.
Alarm
1 beep
every 20
sec.
Alarm
1 beep
every 20
sec.
Alarm
1 beep
every 30
sec.
More Details
Unit will alarm when
battery temperature is
approaching safety
shutdown threshold.
If shutdown threshold
is reached, unit will
shut down. If external
power is connected,
alarm screen will
display and audible
alarm will continue for
2 min
Unit will alarm when
internal temperature is
approaching shutdown
threshold.
If shutdown threshold
is reached, unit will
shut down and alarm
screen will display and
audible alarm will
continue for 2 min
If shutdown threshold
is reached, unit will
shut down and alarm
screen will display and
audible alarm will
continue for 2 min
No breath detected for
15 seconds.
Activated after first
breath or warm up
screen is off,
whichever comes
sooner.
During the No Breath
Detect alarm, the unit
will auto-dose at 20
bpm until a breath is
detected and the alarm
is disabled.
(This feature is only
available for Firmware
Rev 1.2 or higher)
25