Alarm Conditions

Priority # Alarm
Description
1
System
Malfunction
Battery Over
Temp
2
System
Malfunction
High
Temperature
3
System
Malfunction
High
Pressure
4
No Breath
detected
5
Low Battery
Shut Down
6
Warm Up
7
Low
Concentra-
tion
Display
FreeStyle Comfort
Audible
More Details
Signal
1 beep
Unit will alarm when battery tempera-
every 20
ture is approaching saftey shutdown
sec
threshold.
If shutdown threshold is reached, unit
will shut down. If external power is
connected, alarm screen will display and
audible alarm will continue for 2 min.
1 beep
Unit will alarm when internal tempera-
every 20
ture is approaching shutdown threshold.
sec
If shutdown threshold is reached, unit
will shut down and alarm screen will
display and audible alarm will continue
for 2 min.
1 beep
If shutdown threshold is reached, unit
every 20
will shut down and alarm screen will
sec
display and audible alarm will continue
for 2 min
1 beep
No breath detected for 15 seconds.
every 30
Activated after first breath or warm up
sec
screen is off, whichever comes sooner.
1 beep
Alarms until battery runs out of power
every 30
and unit shuts down
sec
none
Unit will start by auto-dosing until a
breath is detected or Warm up ends,
whichever is sooner.
1 beep
every 30
sec, for
up to 5
min
PN MN232 C | User Manual