FreeStyle Comfort
0x0200
0x0400
0x0800
0x1000
0x2000
0x4000
30
Breath Rate
Patient
Exceeded
Breathing more
than 40 BPM
Other
Fan Failure
Other
Low
Intake Filter
Pressure
Clogged
Pinched or
Clogged Inlet
Tubing
Feed/Waste
Valve Failure
Sieve Bed
Failure
Other
Low
Case
Temperature
Temperature
less than 0 °C
Compressor
Other
Failure
O
Out of
Other
2
Range
Run unit in normal operation
for 15 minutes without
breathing off the unit.
Ensure alarm does not occur
during this period.
Contact Caire to send the
unit in for repair.
Contact Caire to send the
unit in for repair.
Clean or replace intake
filters.
Verify that compressor inlet
tubing is not pinched or
clogged.
Replace feed/waste valves.
Check for leaks on bottom
sieve bed caps. Replace
sieve beds.
Contact Caire to send the
unit in for repair.
Allow unit to warm up to
room temperature
Contact Caire to send the
unit in for repair.
Contact Caire to send the
unit in for repair.