topic is necessary. The organisms of most
concern are M. tuberculosis, HIV, and Viral
Hepatitis. These are potentially pathogenic.
Tuberculosis can survive outside the human body,
but its mode of transmission is by droplet nuclei.
When infected individuals a cough, they release
droplet nuclei into the air, and these carry the
Tuberculosis organism. These droplet nuclei may
be breathed in by another person, but prolonged
exposure to the infected person is usually
necessary for infection to occur.
HIV and Viral Hepatitis are both viruses that are
not living cells themselves but can duplicate when
in a living "host" cell. Both of these organisms are
usually passed on by person-to-person contact,
and both need to be in the human body to survive.
Once outside the body, viruses do not survive.
CLEANING AND PREPARING FOR
NEW PATIENT USE
When you remove the FreeStyle Comfort from a
patient's home, always detach and dispose of the
used nasal cannula. Clean the exterior of the unit
with a mild household cleaner applied with a
damp cloth or sponge. Pay special attention to the
oxygen outlet for the cannula connection to make
sure it remains free of dust, water, and particles.
Be careful not to get any liquid into the interior of
the unit.
Next, clean the exterior with a common chemical
disinfectant before any other patient uses the unit.
Do not use liquid directly on the unit to clean it.
WARNING: TURN OFF THE UNIT,
DISCONNECT THE POWER CORD FROM
THE ELECTRICAL OUTLET BEFORE YOU
CLEAN THE UNIT TO PREVENT
ACCIDENTAL ELECTRICAL SHOCK AND
BURN HAZARD
REMOVE THE BATTERY FROM THE UNIT BEFORE
PERFORMING ANY MAINTENANCE TO PREVENT
ACCIDENTAL ELECTRICAL SHOCK AND BURN
HAZARD. THE FREESTYLE COMFORT MUST BE
OFF AND DISCONNECTED FROM XTERNAL
POWER BEOFRE REMOVING THE BATTERY PACK
Clean or replace the air intake filter with warm
soapy water between each patient use. The patients
should be instructed to clean this filter at least once
22
per week, depending on the environment, during
normal operation.
Allow the unit to air dry. Then re-test oxygen
concentration, test the duration of the battery pack,
and ensure the AC and DC power supplies are in
good working condition before you return the unit
to your inventory.
CAUTION: DO NOT USE LIQUID
COMFORT UNIT TO CLEAN IT. A
LIST OF UNDESIRABLE CHEMICAL
AGENTS INCLUDES BUT IS NOT
LIMITED TO THE FOLLOWING,
ACCORDING TO THE PLASTICS
MANUFACTURER: ALCOHOL AND
ALCOHOL-BASED PRODUCTS,
CHLORINE-BASED PRODUCTS
(ETHYLENE CHLORIDE), AND OIL-
BASED PRODUCTS (PINE-SOL,
LESTOIL). THESE ARE NOT TO BE
USED TO CLEAN THE PLASTIC
HOUSING ON THE FREESTYLE
COMFORT, THEY CAN DAMAGE
THE UNIT'S PLASTIC.
Note: Make the bleach solution a 1:100 dilution
of 5.25% sodium hypochlorite. Mix one part
household bleach (e.g., Clorox with 99 parts
cold tap water). To measure the solution easily,
take ¼ cup of household bleach, and mix it with
a gallon of cold tap water. Allow the mixture to
sit on potentially contaminated surfaces for ten
minutes
Note: CAIRE does not recommend the
sterilization of this equipment.